ISO 5360-2016 pdf free.Anaesthetic vaporizers – Agent- specific filling systems.
ISO 5360 specifies requirements. including dimensions. for agent-specific filling
systems br agent-specific anaesthetic vaporizers.
This International Standard does not specify construction materials.
NOTE 1 For recommendations on materials, see Annex A.
Because of the unique properties of desfiurane, dimensions for this agent have not been specified In this International Standard.
NOTE 2 Designs of connection systems. which only permit engagement of the agent-specific bottle adaptor to the bottle when the bottle collar Is in plae. are encouraged,
2 Normative references
The following documents, In whole or in part, are normatively referenced in this document and are Indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (Including any amendments) applies.
ISO 1101, Geometrical Product Specifications (GPS) — Geometrical tokruncirig — Tolerances of form. orientation, location and run-out
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1 agent-sped tic
having both a prescribed configuration and prescribed dimensions, which are specific for a prescribed liquid anaesthetic agent
3.2 agent-specific filling system
functional system of ajent-specafic (11) coded connections between an anaesthetic bottle and an agent- specific (31) anaesthetic vaporizer (13). consisting of, for example, a threaded bottle neck (31) with coibr, bottle connector (16). male adaptor (19). and filler receptacle (lii)
Note ito entry: Different types of agent-specific telling systems are shown in Annex II.
3-3 anaesthetic vaporizer
device designed to facilitate the change of an anaesthetic agent from a liquid to a vapour
4 Bottle
Each bottle shall have the following:
a) name of the anaesthetic agent with which it is intended to be used marked on it;
b) either a bottle collar complying with Clause 5 and a threaded neck complying with Figure 1 and Table 1 or a permanently attached bottle adaptor complying with 6.2.
5 Bottle collar
5.1 Bottle collars shall comply with the configuration and dimensions shown in Figure 2 and angle, 0, specified in Table 2 for the anaesthetic agent with which it is intended to be used.
5.2 The position of the bottle collar relative to the screw thread of the bottle shall be as shown in Figure 3.
5.3 The bottle collar shall be attached to the bottle and shall be rotatable by hand.ISO 5360 pdf download.